Triggering the Oscilloscope
< qualifier time set softkey
> qualifier time set softkey
104
For example, for a positive pulse, if you set t>10 ns:
10 ns
•
within a range of time values (><).
For example, for a positive pulse, if you set t>10 ns and
t<15 ns:
10 ns
5 Select the qualifier time set softkey (< or >), then rotate the
Entry knob to set the pulse width qualifier time.
The qualifiers can be set as follows:
•
2 ns to 10 s for > or < qualifier (5 ns to 10 s for 100 MHz
and 300 MHz bandwidth models)
•
10 ns to 10 s for >< qualifier, with minimum difference of
5 ns between upper and lower settings
• When the less than (<) qualifier is selected, the Entry knob
sets the oscilloscope to trigger on a pulse width less than the
time value displayed on the softkey.
• When the time range (><) qualifier is selected, the Entry
knob sets the upper time range value.
• When the greater than (>) qualifier is selected, the Entry
knob sets the oscilloscope to trigger on a pulse width greater
than the time value displayed on the softkey.
• When the time range (><) qualifier is selected, the Entry
knob sets the lower time range value.
10 ns
15 ns
6000 Series Oscilloscope User's Guide
Trigger
12 ns