Procedure - HP 37717C Calibration Manual

Communications performance analyzer
Hide thumbs Also See for 37717C:
Table of Contents

Advertisement

C AU T I O N
Performance Tests
Transmitted Jitter Amplitude Accuracy (Option UHK)
3. Make the following key sequence on
the HP 37717C to obtain the special
display.
MODULE DEBUG
Press
;
;
OTHER
MORE
;
;
MORE
MORE
Press
until
MORE
MODULE DEBUG
appears in the softkey menu.
Press
MODULE DEBUG
MODULE PDH Module.
Set the VCO Control Mode to FIXED.
The sequence above must be performed each time the power is cycled on the HP
37717C, as all
MODULE DEBUG
is cycled.

Procedure

Transmitted Jitter Amplitude Accuracy on Range 1
1. Setup the HP 37717C
display as shown opposite.
;
;
;
OTHER
. .
and select
parameters adopt DEFAULT values when power
TRANSMIT
3-127