Performance Tests
PDH Receiver Equalization (Options UKK, [USB], UKJ, [USA], UKL,
[USC], UKN, [USE])
6. Press
and set up the display
RESULTS
as shown opposite.
7. Set the Synthesizer to: FREQUENCY -
352 KHz sinewave AMPLITUDE - 2.37 V
pk-pk (6dB down) as measured on the
oscilloscope.
8. Press
RUN/STOP
errors are displayed at the end of the test
period (5 seconds).
2.048 Mb/s
9. Press
and set up the display
RECEIVE
as shown opposite.
10. Select
RESULTS
11. Press
and set up the display
RESULTS
as shown opposite.
12. Set the Synthesizer to: FREQUENCY,
1024 KHz sinewave AMPLITUDE, 2.37
V pk-pk (6dB down) as measured on the
oscilloscope.
13. Press
RUN/STOP
errors are displayed on the
display at the end of the test period (5
seconds).
and verify that no
TIMING
CONTROL
and verify that no
RESULTS
TEST TIMING [SINGLE] [5 SECS]
3-45