Calibration Procedure - HP 54501A Operation Reference

Front panel operation reference digitizing oscilloscope
Hide thumbs Also See for 54501A:
Table of Contents

Advertisement

Calibration
Procedure
Utility Menu
12-12
The process to calibrate the
initiated the process the oscilloscope will display instruetions for
completion.
Before the H P 5450lA can be calibrated the CAllBRATION
toggle switch on the rear panel must be set to UNPROTECTED.
The procedure for calibration is:
Perform a key-down power up to set the
point. This allows the oscilloscope to begin the calibration routines
from the same point each time.
Set the default calibration factors in the Service menu.
Exit the Service menu and enter the self cal menu.
Select 0 in the cal select field and calibrate the vertical factors. This
rouline will calibrate the AID, vertical gain, offset, and trigg e r.
Select 1 in the cal select field and calibrate the delay.
When the software calibrations
toggle switch on the rear panel is set to
CALIBRATION
PROTEcrED.
HP
5450lA is very simple. When you have
Note
HP
5450lA to a known
are
complete reset the