Performance Tests
PDH Binary Interfaces (Option UH3, [US7])
5. Press
PDH
RECEIVE
and set up the display as
SETTINGS
shown opposite.
6. Press
PDH BINARY
RECEIVE
set up the display as shown opposite.
7. Recall default settings on the Digital Transmission Analyzer and select: Interface
- NRZ BIN ECL Frequency - 170 MHz Pattern - 2
8. Check that the HP 37717C Pattern Loss and Signal Loss indicators are not lit.
9. Select the following on the Digital Transmission Analyzer:
Frequency - 700 kHz
15
Pattern 2
-1
10. Select SIGNAL 2 Mb/s and PATTERN 2
displays.
RECEIVE
11. Check that the HP 37717C Pattern Loss and Signal Loss indicators are not lit.
12. Select Interface - NRZ BIN TTL on the Digital Transmission Analyzer and
THRESHOLDS EXT CLOCK - TTL on the HP 37717C
display.
13. Check that the HP 37717C Pattern Loss and Signal Loss indicators are not lit.
MAIN
and
23
15
-1 on the HP 37717C
-1
TRANSMIT
BINARY
TRANSMIT
and
3-171