Performance Tests
PDH Receiver Monitor Levels (Options UKK, [USB])

2.048 Mb/s

9. Select SIGNAL [2 Mb/s] on the
10. Set the Synthesizer to: FREQUENCY - 1024 KHz sinewave AMPLITUDE - 150
mV pk-pk (30dB down) as measured on the oscilloscope.
11. Press
RUN/STOP
period (5 seconds).
8.448 Mb/s
12. Select SIGNAL [8 Mb/s] on the
13. Set the Synthesizer to: FREQUENCY - 4224 KHz sinewave AMPLITUDE - 150
mV pk-pk (30dB down) as measured on the oscilloscope.
14. Press
RUN/STOP
period (5 seconds).
34.368 Mb/s
15. Select SIGNAL [34 Mb/s] on the
16. Set the Synthesizer to: FREQUENCY - 17184 KHz sinewave AMPLITUDE -
100 mV pk-pk (26dB down) as measured on the oscilloscope.
17. Press
RUN/STOP
period (5 seconds).
139.264 Mb/s
18. Select SIGNAL [140 Mb/s] on the
19. Set the Synthesizer to: FREQUENCY - 69632 KHz sinewave AMPLITUDE - 50
mV pk-pk (26dB down) as measured on the oscilloscope.
20. Press
RUN/STOP
period (5 seconds).
RECEIVE
and verify that no errors are displayed at the end of the test
RECEIVE
and verify that no errors are displayed at the end of the test
RECEIVE
and verify that no errors are displayed at the end of the test
RECEIVE
and verify that no errors are displayed at the end of the test
display.
display.
display.
display.