Description; Equipment Required; Procedure - HP 37717C Calibration Manual

Communications performance analyzer
Hide thumbs Also See for 37717C:
Table of Contents

Advertisement

N OTE
N OTE
Performance Tests
PDH Transmitter Output (UKK, [USB], UKJ, [USA]

Description

This test verifies that the PDH transmitter output level and pulse shape meet
required ITU specifications for all rates. The SIGNAL OUT Port is connected to a
Digitizing Oscilloscope and the waveform is checked for required amplitude and
duty cycle at the nominal mid-points. The waveform shape is also checked by
comparison with the special ITU masks. These may be obtained from the
oscilloscope memory (if this feature is fitted) or can be traced from the attached
figures and compared with a printout of the oscilloscope waveform.
A Balanced to unbalanced Converter is required to test the Balanced output port. The
3V peak at the output is reduced to 2.37V peak on the oscilloscope by this device.
The oscilloscope must be terminated in 75
outputs.

Equipment Required

Oscilloscope
120W/75W Balanced to
Unbalanced Converter
ThinkJet Printer

Procedure

704 kb/s Unbalanced Output
1. Recall the HP 37717C DEFAULT SETTINGS as shown on Page 3-2.
2. Connect the Unbalanced 75
a 75
termination and "T" connector. Set the oscilloscope termination to 1M .
Steps 3 to 10 are only valid for Option UKK, [USB].
3-10
for both balanced and unbalanced
: HP 54503A
: HP 15508C
: HP 2225A
SIGNAL OUT port to the Oscilloscope Input 1 via