Pdh Receiver Monitor Levels (Options Ukk, [Usb]); Specification; Description; Equipment Required - HP 37717C Calibration Manual

Communications performance analyzer
Hide thumbs Also See for 37717C:
Table of Contents

Advertisement

Performance Tests

PDH Receiver Monitor Levels (Options UKK, [USB])

PDH Receiver Monitor Levels (Options UKK, [USB])
Specification
Bit Rate
704 kb/s
2.048 Mb/s
8.448 Mb/s
34.368 Mb/s
139.264 Mb/s

Description

The PDH receiver Monitor levels are checked by attenuating the synthesizer output
level by an amount equivalent to the extra gain provided by the Monitor Amplifier.
The synthesizer is set to generate a sinewave at half the bit rate which corresponds to
a ternary all ones signal. The receiver should sync up with no errors if PATTERN
[ALL ONES] is selected.

Equipment Required

Synthesizer
: HP 3335A option 001
Oscilloscope
: HP 54503A

Procedure

1. Recall the HP 37717C DEFAULT SETTINGS as shown on 3-2.
2. Connect up the equipment as shown in Figure 3-22.
3. Set the oscilloscope to: TERMINATION - 1 M
4. Select
RESULTS
Nominal Loss
26 to 30dB
26 to 30dB
26 to 30dB
26dB
26dB
TEST TIMING [SINGLE] [5 SECS]
TIMING
CONTROL
PROBE - 1:1
3-47

Advertisement

Table of Contents
loading