Performance Tests
Jitter Receiver STM-1/STM-4 Optical Interface (Options A1P, [A1S], A3N
[A3P])
9. Press
select STORED SETTINGS NUMBER [0] (Default Settings) and
OTHER
press
.
RECALL
10. Verify that the LASER ON led is off before continuing.
11. Disconnect the Optical Attenuator Output from the HP 8153A and connect to the
Optical In Port on the STM Jitter Receiver Module.
12. Press
TRANSMIT
13. Press
and select SDH Jitter. Set signal to [STM-1 OPT].
RECEIVE
14. Press
TROUBLE SCAN
RESULTS
measurement.
15. After 5 minutes check that NO TROUBLE is displayed on the
display.
16. Press
OTHER
select STORED SETTINGS NUMBER [0] (Default Settings) and
press
.
RECALL
17. Verify that the LASER ON led is off before continuing.
18. Repeat steps 1 through 16, selecting SIGNAL [STM-4 OPT] on both the Transmit
and Receive pages.
19. Disconnect the test equipment.
and select SIGNAL [STM-1 OPT].
SDH
and then
to start a
RUN/STOP
RESULTS
3-167