Figure 3-1
Performance Tests
PDH Transmitter Output (UKK, [USB], UKJ, [USA]
3. Press
TRANSMIT
display as shown opposite.
4. Press
AUTOSCALE
oscilloscope.
5. Adjust the Oscilloscope Timebase and
Delay to position the positive peak pulse
amplitude at mid-pulse-width point in the
centre of the screen.
6. Verify the Pulse in Figure 3-1 meets the following criteria:
Pulse Amplitude
2.133V to 2.607V
Pulse Width
639 ns to 781 ns
Overshoot
0.474V
Undershoot
0.474V
704 kb/s Pulse Criteria
7. Adjust the Oscilloscope Delay to position the negative peak pulse amplitude at
mid-pulse-width point in the centre of the screen and verify that the negative pulse
meets the criteria listed in step 6.

704 kb/s Balanced Output

8. Connect the Balanced to Unbalanced Converter between the HP 37717C 120
Balanced SIGNAL OUT port and the Oscilloscope, leaving the 75
"T" connector in place.
and set up the
on the
termination and
3-11