Performance Tests
Received Jitter Accuracy (Options UHN, [US9], A1M [A1Q], A1N [A1R],
A1P [A1S], A3L [A3M], A3V [A3W], A3N [A3P])
12. Select
TRANSMIT
setup the display as shown below.
13. Select
RESULTS
14. Set up the HP 37717C
shown opposite.
Press
to start the
RUN/STOP
measurement.
When the measurement is complete check
the displayed Hit Count is between
211,860 and 216,140.
Demodulated Jitter Output
15. Connect the HP 37717C Demodulated Jitter Output to the oscilloscope
terminated in 75 .
16. Check that the amplitude of the 10.7 kHz displayed waveform is between 435mV
pk_pk and 565mV pk_pk.
17. Disconnect all test equipment.
3-152
and
JITTER
TEST TIMING [SINGLE] [10 SECS].
TIMING CONTROL
as
RESULTS