Performance Tests
PDH Receiver Equalization (Options UKK, [USB], UKJ, [USA], UKL,
[USC], UKN, [USE])

8.448 Mb/s

14. Select SIGNAL [8 Mb/s] on the
15. Set the Synthesizer to: FREQUENCY - 4224 KHz sinewave AMPLITUDE - 2.37
V pk-pk (6dB down) as measured on the oscilloscope.
16. Press
RUN/STOP
display at the end of the test period (5 seconds).
34.368 Mb/s
17. Select SIGNAL [34 Mb/s] on the
18. Set the Synthesizer to: FREQUENCY - 17184 KHz sinewave AMPLITUDE - 0.5
V pk-pk (12dB down) as measured on the oscilloscope.
19. Press
RUN/STOP
display at the end of the test period (5 seconds).
139.264 Mb/s
20. Select SIGNAL [140 Mb/s] on the
21. Set the Synthesizer to: FREQUENCY - 69632 KHz sinewave AMPLITUDE -
0.25 V pk-pk (12dB down) as measured on the oscilloscope.
22. Press
RUN/STOP
display at the end of the test period (5 seconds).
3-46
RECEIVE
and verify that no errors are displayed on the
RECEIVE
and verify that no errors are displayed on the
RECEIVE
and verify that no errors are displayed on the
display.
RESULTS
display.
RESULTS
display.
RESULTS