Performance Tests
Wander/Slips Measurement (Options UHN, [US9], A1M [A1Q], A1N
[A1R], A1P [A1S], A3L [A3M], A3V [A3W], A3N [A3P])
3. Press
and set up the
RECEIVE
display as shown opposite.
4. Press
and set up the display
RESULTS
as shown opposite.
5. Connect the unbalanced 75
OUT port to the unbalanced 75
port and the unbalanced 75
REF INPUT (Jitter Receiver Module)
simultaneously using the T-piece.
6. Press
RUN/STOP
display shows POSITIVE PEAK
WANDER and NEGATIVE PEAK
WANDER readings of 0.000 0.125 BITS.
7. Press
RUN/STOP
8. Disconnect the PDH OUT port from the unbalanced 75
and connect to the balanced 120
Balanced to Unbalanced Converter.
9. Press
and select WANDER REFERENCE [120
RECEIVE
10. Press
RUN/STOP
WANDER and NEGATIVE PEAK WANDER readings of 0.000 0.125 BITS.
11. Remove the input from the TIMING REF port and check that the display
indicates NO REF.
12. Press
RUN/STOP
PDH
PDH IN
TIMING
and verify that the
to stop the measurement.
TIMING REF INPUT via the HP 15508C
and verify that the display shows POSITIVE PEAK
to stop the measurement.
TIMING REF INPUT
BAL].
3-155