HP 37717C Calibration Manual page 215

Communications performance analyzer
Hide thumbs Also See for 37717C:
Table of Contents

Advertisement

Performance Tests
STM-1/STM-4, 1550nm Optical Interface (Option URU)
2. Setup the HP 8153A as follows:
a. Press
key to display wavelength [ ]
PARAM
b. Using
,
c. Press
key to display Time [t]
PARAM
d. Using
,
e. Press
key to display REF.
PARAM
f. Using
,
g. Press
key to display CAL.
PARAM
h. Using
,
i. Cover the end of the Power Meter Optical Input connector with a protective cap.
j. Press the
ZERO
ready.
3. Switch on the HP 37717C, check that immediately on power-up the Optical
Module LASER ON led illuminates for a few seconds and Recall the HP 37717C
DEFAULT SETTINGS as shown on 3-2.
4. Press
TRANSMIT
37717C and set up the display as shown
opposite.
5. Verify that the Optical Module LASER
ON led is lit indicating that the laser is
enabled.
6. Adjust the level to obtain a reading of-
8dBm on the Power Meter. Typically
around 6dB of attenuation will be required
to achieve this reading.
7. Press
TRANSMIT
8. Verify that the Optical Module LASER ON led is NOT lit indicating that the laser
is disabled before continuing.
9. Disconnect the optical cable from the Power Meter and connect to the HP 37717C
STM1/4 Optical Rx Input.
3-106
and
keys, set the wavelength to 1550nm.
and
keys, set the time to 200mS.
and
keys, set the REF to 0.000dBm.
and
keys, set the CAL to 0.000dBm.
key on the Power Meter to calibrate - the Power Meter is now
on the HP
SDH
on the HP 37717C and select SIGNAL [STM-1].
SDH