78
Repair procedures

Replacing Recharger Board fuse (F2)

The Recharger Board fuse (F2) protects the monitor power supply input at the battery line
to the monitor. If the monitor does not turn on but the BATTERY CHARGING indicator
lights when the ac power adapter is connected, check fuse F2 (3 Amp).
1.
Remove the battery pack
2. Open the monitor
3. Remove the Recharger Board
4. Replace the fuse (F2, shown below).
fuse F2fuse F2
Caution The fuse is heat sensitive and must be soldered with heat sinks in
place between the fuse body and the solder joint.
("Removing the single battery pack"
("Opening the monitor"
on page 68).
("Replacing the Recharger Board"
F2
P2
T3
D33
Welch Allyn Propaq CS Vital Signs Monitor
on page 66).
on page 83).