Welch Allyn PROPAQ CS Service Manual page 30

Vital signs monitor
Hide thumbs Also See for PROPAQ CS:
Table of Contents

Advertisement

26
Functional verification
nibp test screen
NIBP TEST
PR
PR
DSC
LEAK
BLEED
7 .
Manually reduce the pressure as indicated on the pressure meter to each of the levels
shown in the following table, and check the displayed value of PR1 to be within the
tolerance listed in the following table.
Pressure (mm Hg)
8. Reduce the pressure to 0 mm Hg and check that the displayed readings (PR1 and
PR2) are 0 ±2 mm Hg.
9. Close the bulb valve.
10. Press CANCEL. (If there is no CANCEL key at this time, go to the next step.)
11. Disconnect the adult cuff from the test setup and connect the neonatal cuff #3 that is
wrapped around a solid 3 cm diameter cylinder.
12. Press PUMP TEST.
13. Check that the pump inflates the cuff to approximately 280 mmHg.
14. After four minutes, check that the displayed leakage is less than 50 mmHg.
Note
15. Press HOME
16. Disconnect the cuff and Cuff Calibration Kit.
ADULT
10:42:17
BATTERY: 8.3 VOLTS
CNTS
mmHg
578
249.4
1
259
338
2
1600
10
mmHg
mmHg/s
PUMP
AUTO
CANCEL
TEST
PUMP
Tolerance (mm Hg)
270
250
200
100
50
20
If the leakage exceeds this specification, first check all exterior hoses and fittings
before diagnosing an internal air leak problem.
.
PRINT FAULT
DAC
VALVE
130
0
0
DIAG:
PUMP:
secs
ZERO
PREVIOUS
±
4
±
3
±
2
±
2
±
2
±
2