138
Fluke patient simulator modification
Incompatibilities with other simulators may also cause erroneous readings and/or missing
ECG. Check Propaq CS monitor readings with the simulator against a known source
before using the simulator for calibration.
Function
Output (+)
Output (-)
Exciter (+)
Exciter (-)
ECG Ref
Color
Pin No.
Propaq CS 244/246
Black
4
C
Red
1
A & E
White
3
B
Green
5
D
Blue
2
No Connection
Welch Allyn Propaq CS Vital Signs Monitor
W1
W2
810-0696-57