Directions for Use
9. Replace the battery door and screws. Turn the electrocardiograph rightside-up.
It turns on automatically and displays a prompt to enter date and time.
10. Enter the date and time.
The electrocardiograph is ready to use.
11. Discard the old battery appropriately.

Replacing the Battery (DC) Fuse

If the battery (DC) fuse requires replacing, as described in
steps. For the fuse value, see
1.
Remove and discard the fuse. See
You may need to pull hard. Use pliers gently if necessary.
2. Connect a new fuse. It goes in either way.
3. Go to
Figure 55. Battery Fuse Removed
In the USA, call 1-800-SAV-LEAD for instructions on how to recycle it.
International users, contact your local authorities concerning recycling.
"Fuses"
Step 8
on page 84.
Chapter 7 Maintaining the Electrocardiograph
Step 5
on page 95.
Figure
55.
85
on page 83, follow these