x = 7
yyy-zzzz

Battery capacity check

Several variables affect monitor operating time on battery:
Refer to the Propaq CS Directions For Use for battery care information and expected
operating times. If the monitor fails to operate on battery for an extended period of time,
follow the procedure below to check the capacity of the monitor's lead acid battery. Lead
acid batteries have a very limited shelf life; you must test the battery if the monitor has
been stored for two or more months without the ac power adapter plugged in.
1.
Use the ac power adapter and charge the monitor for at least 8 hours with the
monitor turned off (12 hours if a printer or pulse oximeter is attached).
2. Disconnect the cuff and all cables from the monitor.
3. Turn off all printing (if a printer is attached).
4. Disconnect the power adapter.
5. Turn on the monitor.
6. Run the monitor for the typical operating time as shown in the table. Confirm that the
monitor does not shut off before the specified time.
Monitor configuration
Stand-alone monitor
Monitor with SpO
(no Expansion Module)
Monitor with Expansion Module with
printer, SpO
7 .
Turn off the monitor.
8. Use the ac power adapter and charge the monitor for at least 8 hours with the
monitor turned off. (Charge at least 12 hours if a printer or pulse oximeter is attached.)
indicates a problem found by the Side Stream Processor (SSP) on the
sidestream main board.
indicate specific types of problems. Contact Welch Allyn Technical
Support for more details.
active options
frequency of NIBP measurements
frequency and length of printouts
ambient temperature,
battery age and condition
displayed information.
Caution A new battery should pass the following test. The run time of older
batteries will decrease proportionally with age. Replacement is recommended
when the run time becomes insufficient for the monitor's intended application.
option
2
, and CO
options
2
2
Typical operating time with a new battery
2 hours
4 hours
3 hours
Troubleshooting