44
Calibration

Equipment needed

You will need the following equipment to calibrate the Propaq CS monitor. Part numbers
are given for equipment available from Welch Allyn.
Note
Table 11. Required equipment
Description
Digital multimeter (DMM), 4 1/2 digit, Fluke 805A or equivalent
50MHz triggered sweep oscilloscope, Tektronix 2225 or equivalent
Variable dc power supply, 0-28 V, 3 A, VIZ WP-715A or equivalent
Patient simulator, Dynatech/Nevada 213A, 214A, 215A, or 217A with temperature and EGG cable/leads
0 to 300mmHg digital manometer with bulb (Netech, Digamano, or equivalent)
Invasive blood pressure (IBP) simulator, 5μ/V/mmHg, Fogg Systems BP48C, BP28, or MDE Datasim 6000, with IBP cables
Adult cuff (provided with monitor); cylindrical object about the size of an adult's arm
Welch Allyn cuff calibration kit, Welch Allyn part number 008-0012-XX
Propaq CS ac power adapter (provided with monitor)
Resistor test loads (4.9Ω and 4.65Ω). Refer to
DC supply cable, Welch Allyn part number 008-0290-XX
Tools necessary to disassemble and reassemble the monitor. (See
tools and instructions on opening the monitor.)

Setup

To calibrate the monitor you must open and then perform some disassembly to reach the
necessary calibration points.
1.
Before you begin disassembly, charge the battery pack for at least 8 hours to ensure it
is fully charged.
2. Remove any options attached to the monitor. If either the SpO
is attached to the monitor, remove it before starting the calibration. The calibration
procedures in this section are for the three monitor boards only.
3. Remove the battery pack.
4. Disconnect the ac power adapter from the monitor's right side panel power input
connector.
All test equipment must be calibrated to function within parameters specified by
the manufacturer, and be traceable to a national calibration standard.
Caution All calibration, service, and repair must be done at a static-protected
work station. Failure to protect against damage due to static discharge can
damage components in the Propaq CS monitor.
Caution Refer to
"Repair procedures"
reassembly procedures. Failure to follow disassembly procedures may result in
damaged internal parts.