Welch Allyn PROPAQ CS Service Manual page 16

Vital signs monitor
Hide thumbs Also See for PROPAQ CS:
Table of Contents

Advertisement

12
Overview
Network Monitor screen (2.4 GHz radio card)
To access this screen, from the Network Status screen press NETORK MONITOR.
(Or from the Main Menu, press SETUP > MORE > MORE > SERVICE > YES > MORE >
MORE > RADIO STATUS > NETWORK MONITOR.)
Network Status screen (5 GHz radio card)
To access this screen, from the Main Menu press SETUP > MORE > MORE > SERVICE >
YES > MORE > MORE > RADIO.
0
Note
Note
NETWORK MONITOR
TX PACKETS:
RX PACKETS:
TX RETRIES:
RX CRC ERRS:
LOAD LEVELING:
AP DROPPED MUS:
NETWORK
STATUS
NETWORK STATUS
FIRMWARE VER:
1.00.00 A0023
NET NAME:
pco|99
MONITOR MAC:
00:1A:FA:00:01:02
MONITOR IP:
172.29.51.32
Before the monitor has made a connection to the network, the fields in the above
screen will not be populated. Once the fields are populated, they will not be
updated until the monitor's power is cycled.
There is no 5 GHz Network Monitor service screen.
BATTERY: 9.2 VOLTS
DIR
UNDIR
ERRORS
58833
813
49431
4604
1954
1MBPS
2MBPS
PERCENT
17065
34997
1714
240
LOW RSSI:
2
3
BATTERY: 8.4 VOLTS
Welch Allyn Propaq CS Vital Signs Monitor
25
S
(M)
9
D
( 15)
mmHg
122
NIBP
S
58
(M)
0
D
( 85)
mmHg
TEMP
100.4
T1
98.6
T2
1.8
ΔT
F
4
MCO
BR
2
2
38
12
mmHg
Br/m
1
S
p
O
2
97
%
PREVIOUS
MENU
T1
37.0
°C
PREVIOUS
MENU

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents