Replacing cables

If you suspect a cable is faulty, check its continuity on all pins. If you need to replace a
cable, refer to
for cable locations.
When removing a cable, be sure to note the routing. Be sure to route the new cable in the
same manner.
P10
P5

Replacing PROMs

Please read the information in
The PROMs require a special removal tool to replace them. See
page 60. To prevent damage to the PROM sockets, do not use any other tools.
The following instructions describe how to replace PROMs:
Note
To remove the PROMs, insert the fingers of the PROM removal tool into the slots at
two corners of the PROM and pull out the PROM, while squeezing the legs against
the PROM.
To load the new PROM, align the beveled corner of the PROM with beveled OUTSIDE
corner of the socket. Carefully press the new PROM into place.
"Field Replaceable Units (FRUs)"
P6
P7
"Propaq CS Monitor software"
Caution The PROMs can be damaged by static discharge. Work only at a static-
protected work station.
Do not remove the PROMs from the Digital Board. For assistance in servicing the
Digital Board and software, contact Welch Allyn or your local distributor.