Functional verification
Required equipment
You will need the following equipment to verify that all functions of the monitor operate
correctly. All test equipment must be calibrated to function within parameters specified
by the manufacturer, and must be traceable to a national calibration standard.
Some equipment can be manufactured. See
page 135 for information on manufacturable test equipment.
Description
Luer connector, male, for use with 1/8" I.D. tubing (Value Plastics, MTLL230-1)
Luer connector, female, for use with 1/8" I.D. tubing (Value Plastics, FTLL230-1)
Test water trap, Welch Allyn, 008-0370-00
1/8" inside diameter tubing
Flow meter, Cole Parmer E 32000-06, 200ml/min. or equivalent
0 to 300 mm Hg calibrated digital pressure meter, with inflation squeeze bulb (Netech, Digamano, or equivalent)
50 MHz triggered sweep oscilloscope
Variable dc power supply, 0-28 V, 3A with voltage and current metering
Patient simulator, Fluke (Dynatech/Nevada) 213A, 214A, 215A, or 217A with IBP, Temperature, and EGG Cable/Leads
Adult cuff (provided with monitor); Neonatal cuff #3; 3 cm diameter solid cylinder; solid cylindrical object about the size
of an adult's arm
Welch Allyn cuff calibration kit, Welch Allyn PN 008-0012-XX
Power supply adapter cable, Welch Allyn PN 008-0290-00
Physio-Control LIFEPAK 5 or LIFEPAK 6s with appropriate Welch Allyn Defib Sync adapter to use as Defib Sync Marker
pulse source
Propaq ac power adapter (provided with the monitor)
Safety analyzer, Fluke (Dynatech/Nevada), 431F-1D or equivalent
Rod-L Model M100AVS5 high-voltage potential (hi-pot) tester, or equivalent
Static-free work area
SpO
Simulator, Fluke (Biotek) Index 2 XL/XLFE
2
Test gas source: dry 4% to 10% CO
concentration must be certified to ±0.01. (Convert percentage to partial pressure (mm Hg) as instructed in
pressure of test gas source"
Two rubber test tube stoppers, with 1/8" to 1/4" holes drilled through (for CO
a.
Of the gas tolerance available, usually described as ±0.03% absolute, the container must contain at least 2,000 psi grams of gas, to
be within ±0.01% of the certified percentage of gas stated on the tag supplied by the vendor. The actual percentage of gas in the
container may vary from lot to lot.
"Manufacturable test equipment"
, balance air (with flow meter). Gas may be between 4% and 10%, but exact CO
2
a
on page 31.)
Welch Allyn Propaq CS Vital Signs Monitor
on
"Partial
tests)
2
2