Performance Tests; Dc Voltage Amplitude Accuracy (Normal); Performing A Calibration Check - Fluke 5500A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

7-6.

Performance Tests

For the following performance tests, refer to Chapter 4, "Front Panel Operations," for
procedures on operating the 5500A from the front panel, or to Chapter 5, "Remote
Operations," for procedures on operating the 5500A from a terminal or computer. Also
refer to Chapter 4 for connection information.
Zero the 5500A Calibrator before testing by completing "Zeroing the Calibrator" in
Chapter 4.
The performance tests have reserved columns for recording the Measured Value and
Deviation (%).
7-7.

DC Voltage Amplitude Accuracy (NORMAL)

The DC Voltage Amplitude Accuracy test verifies the accuracy of dc voltage at the
5500A Calibrator front panel NORMAL terminals.
Range
Nominal Value
330 mV
0.0000 mV
330 mV
329 mV
330 mV
-329 mV
3.3 V
0.000 mV
3.3 V
3.29 V
3.3 V
-3.29 V
33 V
0.00 mV
33 V
32.9 V
33 V
-32.9 V
330 V
50 V
330 V
329 V
330 V
-50 V
330 V
-329 V
1020 V
334 V
1020 V
1020 V
1020 V
-334 V
1020 V
-1020 V
Measured Value
Deviation %
Maintenance

Performing a Calibration Check

90-Day Spec. ( V or %)
3.0 V
0.0059%
0.0059%
5 V
0.0042%
0.0042%
50 V
0.0042%
0.0042%
0.0055%
0.0047%
0.0055%
0.0047%
0.0049%
0.0046%
0.0049%
0.0046%