Impedance/Capacitance Function Commands - Fluke 5500A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

5500A
Operator Manual
TLIMIT_D
(IEEE-488, RS-232, Sequential)
Sets the default OPERATE time limit for the OVERLD mode signal.
Parameters: 1 to 60 (seconds)
Example:
TLIMIT_D?
(IEEE-488, RS-232, Sequential)
Returns the default overload time limit.
Response:

8-55. Impedance/Capacitance Function Commands

ZERO_MEAS
(IEEE-488, RS-232, Sequential)
Sets the measurement offset to the capacitance value.
Parameters: (boolean) ON or OFF.
*TRG
(IEEE-488, RS-232, Sequential)
Triggers and returns a new impedance measurement value when used with the SC600
option in MEAS Z mode. (See Chapter 6 for *TRG use in all cases except MEAS Z
mode with the SC600 option.)
Responses: <measurement value>, OHM
Example:
You can also use the VAL? query to return an impedance measurement
value with the SC600 option. VAL? returns the last measurement, whereas
*TRG gets a new measurement. Responses are the same as shown above
for the *TRG command. (See Chapter 6 for VAL? use with thermocouple
measurements.)
8-40
TLIMIT_D 15
<Integer> Default time limit in seconds.
<measurement value>, F
<measurement value>, NONE
*TRG returns 1.00E+03,OHM
(input impedance value in ohms)
(input capacitance value in farads)
(no measurement is available)
(1 k input impedance).