Leveled Sinewave Function Verification: Amplitude - Fluke 5500A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

8-118. Wave Generator Function Verification: 50
Nominal Value
Waveform
(p-p)
Square
5.0 mV
Square
10.9 mV
Square
44.9 mV
Square
109.0 mV
Square
449.0 mV
Square
1.1 V
Square
2.2 V
Sine
5.0 mV
Sine
10.9 mV
Sine
44.9 mV
Sine
109.0 mV
Sine
449.0 mV
Sine
1.1 V
Sine
2.2 V
Triangle
5.0 mV
Triangle
10.9 mV
Triangle
44.9 mV
Triangle
109.0 mV
Triangle
449.0 mV
Triangle
1.1 V
Triangle
2.2 V

8-119. Leveled Sinewave Function Verification: Amplitude

Nominal Value (p-p)
Frequency
5.0 mV
10.0 mV
20.0 mV
40.0 mV
50.0 mV
100.0 mV
200.0 mV
Measured Value
Frequency
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
10 kHz
Measured Value (p-p)
50 kHz
50 kHz
50 kHz
50 kHz
50 kHz
50 kHz
50 kHz
Load
Deviation
(p-p)
(mV)
Deviation
(mV)
5500A-SC300 Option
Verification Tables
1-Year Spec.
(mV)
0.25 mV
0.43 mV
1.45 mV
3.37 mV
13.57 mV
32.50 mV
66.10 mV
0.25 mV
0.43 mV
1.45 mV
3.37 mV
13.57 mV
32.50 mV
66.10 mV
0.25 mV
0.43 mV
1.45 mV
3.37 mV
13.57 mV
32.50 mV
66.10 mV
1-Year Spec.
(mV)
0.300
0.400
0.600
1.000
1.200
2.200
4.200
8-91
8

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents