Testing A Model 41 Power Harmonics Analyzer; Testing Watts, Va, Var Performance - Fluke 5500A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

5500A
Operator Manual
1. Verify that the calibrator is set to 0 V dc in standby. Press
2. Turn on the 80 Series DMM, and set its function switch to
3. Connect a set of test leads to the DMM as shown in Figure 4-15.
4. Set the calibrator to 3.5 V dc and press
5. The DMM should now display 3.500 0.001. If necessary, adjust R21 to obtain the
6. Set the DMM function switch to
7. The DMM should display 3.500 0.002. If necessary, adjust R34 to obtain the proper
8. Change the calibrator output to 10 kHz.
9. The DMM should display 3.500 0.004. If necessary, adjust C2 to obtain the proper
10. Change the calibrator output to 35.00 V at 10 kHz.
11. The DMM should display 35.00 0.04. If necessary, adjust C3 to obtain the proper

4-58. Testing a Model 41 Power Harmonics Analyzer

The Model 41 Power Harmonics Analyzer, hereafter referred to as the Tester, requires
two voltages at varying phase relationships to test the functionality of the Power and
Harmonics features. The procedure for testing these two functions of the Tester are
included here to demonstrate the operation of the dual voltage function of the Fluke
5500A.
These procedures are included here as an example. The Model 41 Service
Manual contains the complete authoritative testing and calibration
procedures.

4-59. Testing Watts, VA, VAR Performance

Perform the following procedure to test the Watts, VA, and VAR functions of the Tester.
Refer to Table 4-4.
Ensure that the calibrator is in standby mode before making
any connection between the calibrator and Tester. Dangerous
voltages may be present on the leads and connectors.
4-58
proper display.
100 Hz.
display.
display.
display.
.
and set the calibrator output to 3.500 V at
Note
Warning
if it is not.
.