General Specifications

Input Power

Voltage ................................................................... 100 V to 240 V with up to ±10 % fluctuations
Frequency .............................................................. 47 Hz to 63 Hz
Maximum consumption ........................................ 3,500 VA Max

Dimensions

Size
Height.................................................................. 1,170 mm (46 in)
Width ................................................................... 600 mm (24 in)
Depth .................................................................. 800 mm (32 in)
Weight .................................................................... 170 kg (374 lb)
Weight in shipment crate ..................................... 210 kg (462 lb)

Environment

Warm-Up Time ....................................................... 1 hour
Temperature and Performance
Operating ............................................................ 5 °C to 35 °C
Storage ............................................................... 0 °C to 50 °C
Transit temperature range ................................. 0 °C to 50 °C. Transit outside this temperature range (-20 °C to 0 °C,
Relative Humidity
Operating ............................................................ <80 % 5 °C to 30 °C ramping down linearly to 50 % at 35 °C
Storage ............................................................... <95 %, non-condensing 0 °C to 50 °C
Operating Altitude ................................................. 0 m to 2,000 m
Storage Altitude .................................................... 0 m to 12,000 m

Standard and Agency Approvals

Safety ..................................................................... Complies with IEC 61010-1, Overvoltage Category II, Pollution degree
EMI/RFI/EMC .......................................................... IEC 61326-1: Controlled EM Environments, FCC part 15 sub-part B

Electrical Source Accuracy Specifications

See specifications in the 6100B/6105A Users Manual.

Performance Specifications for IEEE C37.118.1a-2014 Tests

Parameter
Steady State
Dynamic, Modulation
Dynamic, Ramp
Parameter
Step Time Accuracy
Measurement reporting latency test accuracy (s)
THD of fundamental (% of setting)
or 50 °C to 60 °C) must be limited to <100 hours
2
class A
TVE (%)
0.1
0.3
0.1
Introduction and Specifications
General Specifications
FE (Hz)
RFE (Hz/s)
0.0005
0.001
0.001
0.02
0.0005
0.01
Delay Time (s)
0.0002
0.00005
0.02
1