1-29. Additional Specifications

The following paragraphs provide additional specifications for the 5500A Calibrator ac voltage and ac current functions.
These specifications are valid after allowing a warm-up period of 30 minutes, or twice the time the 5500A has been turned
off. All extended range specifications are based on performing the internal zero-cal function at weekly intervals, or when
the ambient temperature changes by more than 5 C. (See Chapter 4, Front Panel Operations in the 5500A Operator
Manual.)

1-30. Frequency Specifications

Frequency
Range
0.01 - 119.99 Hz
120.0 - 1199.9 Hz
1.200 - 11.999 kHz
12.00 - 119.99 kHz
120.0 - 1199.9 kHz
1.200 - 2.000 MHz
[1]
± (25 ppm + 15 mHz) above 10 kHz
1-31. Harmonics (2
Fundamental
[1]
Frequency
10 to 45 Hz
45 to 65 Hz
65 to 500 Hz
500 to 1 kHz
1 to 5 kHz
Phase uncertainty for harmonic outputs is 1 degree, or the phase uncertainty shown in "Phase Specifications" for the particular output,
whichever is greater. For example, the phase uncertainty of a 400 Hz fundamental output and 10 kHz harmonic output is 10 degrees
(from "Phase Specifications"). Another example, the phase uncertainty of a 60 Hz fundamental output and a 400 Hz harmonic output is
1 degree.
[1]
The maximum frequency of the harmonic output is 10 kHz. For example, if the fundamental output is 5 kHz, the maximum
selection is the 2
200 Hz.
Example of determining Amplitude Uncertainty in a Dual Output Harmonic Mode
What are the amplitude uncertainties for the following dual outputs?
NORMAL (Fundamental) Output:
100 V, 100 Hz ................................................. From "AC Voltage (Sine Wave) Specifications" the single output
th
AUX (50
Harmonic) Output:
100 mV, 5 kHz ................................................ From "AC Voltage (Sine Wave) Specifications" the auxiliary output
1-26
Resolution
0.01 Hz
0.1 Hz
1.0 Hz
10 Hz
100 Hz
1 kHz
nd
th
to 50
) Specifications
Voltages
NORMAL Terminals
33 mV to 32.9999 V
3.3 mA to 2.19999 A
33 mV to 1020 V
33 mV to 1020 V
330 mV to 1020 V
3.3 to 1020 V
33 mA to 2.19999 A
nd
harmonic (10 kHz). All harmonic frequencies (2
specification for 100 V, 100 Hz, is 0.015 %
in this example, the specification is 0.015 %
the same, and the floor is twice the value (2 x 2 mV).
specification for 100 mV, 5 kHz, is 0.15 %
output in this example, the specification is 0.15 %
0.15 % is the same, and the floor is twice the value (2 x 450 mV).
1-Year Absolute Uncertainty,
tcal ± 5 C
± (PPM + mHz)
25
25
25
25
25
25
Voltages
Currents
AUX Terminals
10 mV to 3.3 V
3.3 mA to 11 A
10 mV to 3.3 V
33 mA to 11 A
100 mV to 3.3 V
33 mA to 11 A
100 mV to 3.3 V
100 mV to 3.3 V
nd
th
to 50
) are available for fundamental outputs between 10 and
Jitter
1
2 s
1
2 s
[1]
1
2 s
15
140 ns
15
140 ns
15
140 ns
Amplitude
Uncertainty
Same % of output as the
equivalent single output,
but twice the floor adder.
2 mV. For the dual output
4 mV as the 0.015 % is
450 mV. For the dual
900 mV as the

Hide quick links: