Performance Tests; Test The Voltage Function - Fluke 150X Series Calibration Manual

Insulation testers
Table of Contents

Advertisement

Performance Tests

The subsequent series of tests comprise a performance test to verify the accuracy of the
Tester (DUT) and its performance level. The performance test is recommended as an
acceptance test for incoming inspection and as a calibration procedure to periodically ensure
the accuracy of the Tester. Fluke recommends that you run the performance test at least once
a year.
No adjustments are required during the performance test, and it is not necessary to open the
case. If the Tester does not pass all parts of the performance tests, repair and/or calibration
adjustment are required. A calibration adjustment procedure is given in
Adjustment. If significant repairs are required, contact Fluke, see
are appropriate, refer to

Test the Voltage Function

To verify the accuracy of the voltage measure function:
1. Turn the DUT rotary switch to the
2. Connect the Fluke 5320A voltage output terminals to the
the DUT.
3. Set the 5320A for the voltage output function (V).
4. Apply the input level for each step listed in
5. Compare the reading on the DUT display with the display reading in
6. If the display reading falls outside of the limits shown in
specification.
Step
Function
1
2
3
4
5
V
6
7
8
9
10
Service and
Parts.
function.
V
Table 4. Voltage Accuracy Tests
Range
3 V, 0 Hz
8 V, 0 Hz
8 V, 400 Hz
50 V, 0 Hz
100 V, 0 Hz
600.0 V
250 V, 0 Hz
500 V, 0 Hz
120 V, 60Hz
230 V, 50 Hz
600 V, 400 Hz
INSULATION
Table
4.
Table
Applied Input
Insulation Testers
Performance Tests
Calibration
Contact
Fluke. If user repairs
and
terminals of
COM
Table
4.
4, the Tester does not meet
Display Reading
Low Limit
High Limit
2.7
3.3
7.5
8.5
Display must show V ac
annunciator
48.7
51.3
97.7
102.3
244.7
255.3
489.7
510.3
117.3
122.7
225.1
234.9
587.7
612.3