4-56. Testing the Meter

You can use the error mode feature of the calibrator to test the meter. To verify that all
ranges of all functions are within specifications, proceed as follows:
1. Turn on the calibrator and allow it to warm up.
2. Verify that the EARTH indicator is on; if not, press
Ensure that the calibrator is in standby mode before making
any connection between the calibrator and tester. Dangerous
voltages may be present on the leads and connectors.
3. Verify that the calibrator is in standby and connect the DMM as shown in
Figure 4-15.
87
TRUE RMS MULTIMETER
MIN MAX
RANGE
HOLD
H
REL
Hz
PEAK MIN MAX
mV
mA
V
A
A
V
OFF
COM
V
mA A
A
Figure 4-15. Cable Connections for Testing an 80 Series General Functions
4. Test the dc voltage function as follows:
a.
Turn on the DMM and set its function switch to
b.
Set the warmed up calibrator to 3.5 V dc. Press
c.
Use the output adjustment controls to adjust the calibrator output for a reading
of +3.5000 on the DMM display.
d.
Verify that the error shown on the control display is less than the specification
for the DMM in its Users Manual.
e.
Check the DMM error at 35.0 V, -35.0 V, 350.0 V. Hint: use the
the errors are within specification. When
33 V, the calibrator goes into standby. When this happens, press
operate.
f.
Check the DMM error at 1000 V to verify it is within specification.
g.
Set the output of the calibrator to 350 mV and press
within specifications.
as necessary.
Warning
.
.
causes the output to go over
Front Panel Operation
Sample Applications
5500A
CALIBRATOR
NORMAL AUX
SCOPE
200V PK
V,
,
A,
-SENSE,
MAX
RTD
AUX V
HI
1000V
20V
RMS
RMS
MAX
MAX
TRIG
LO
OUT
1V PK
MAX
20V PK
MAX
TC
20V PK
MAX
f4-15.eps
. Verify
to
. Verify the errors are
4
4-53

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents