Setting Dc Voltage Output - Fluke 5500A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

5500A
Operator Manual

4-21. Setting DC Voltage Output

Complete the following procedure to set a dc voltage output at the 5500A front panel
NORMAL terminals. If you make an entry error, press
reenter the value.
Verify the applied voltage to the UUT does not exceed the
rating of the UUT insulation.
1. Press
2. Connect the UUT as described earlier in this chapter under "Connecting the
3. Set the UUT to measure dc voltage on the desired range.
4. Press the numeric keys and decimal point key to enter the desired voltage output
At voltage outputs of 100 volts and above (nominal), you may notice a
slight high-pitched sound. This is normal.
5. Press
6. Press a multiplier key, if necessary. For example, press
7. Press
8. The Control Display now shows the amplitude of your entry. For example,
9. Press
10. Press
4-18
Dual AC voltage
Capacitance
Temperature - RTD
Temperature - Thermocouple
Resistance
Boost operation (using an auxiliary amplifier)
to clear any output from the 5500A.
Calibrator to a UUT."
(maximum seven numeric keys). For example, 123.4567.
to select the polarity of the voltage (default is +).
.
123.4567 mV (below).
. The calibrator clears your entry from the Control Display and copies
it into the Output Display (below is typical).
to activate the calibrator output.
Caution
Note
to clear the display, then
.