7-21. AC Power Amplitude Accuracy (High Voltage)

This verification test is optional. It is not necessary to guarantee the full
calibration of the instrument. However, it may be useful when
troubleshooting an instrument when other functions and/or parameters are
incorrect.
The AC Power Amplitude Accuracy (High Voltage) tests the current outputs at the AUX
terminals in the presence of a high voltage.
Nominal
Nominal
Value (V)
Value
(NORMAL)
(A)
(AUX)
1000 V
3.3 mA
1000 V
3.3 mA
1000 V
33 mA
1000 V
33 mA
1000 V
33 mA
1000 V
33 mA
1000 V
33 mA
(Optional)
33 mA
800 V

7-22. AC Power Amplitude Accuracy (High Current)

This verification test is optional. It is not necessary to guarantee the full
calibration of the instrument. However, it may be useful when
troubleshooting an instrument when other functions and/or parameters are
incorrect.
The AC Power Amplitude Accuracy (High Current) tests the voltage outputs at the
NORMAL terminals in the presence of a high current.
Nominal
Nominal
Value (V)
Value
(NORMAL)
(A)
(AUX)
33 mV
11 A
33 mV
11 A
330 mV
11 A
3.3 V
2.19 A
3.3 V
329 mA
Note
Phase
Frequency Hz
(degrees)
65 Hz
0
65 Hz
90
500 Hz
0
500 Hz
90
1 kHz
0
5 kHz
0
7 kHz (10 kHz
0
optional)
10 kHz
Note
Phase
Frequency
(degrees)
Hz
65 Hz
0
65 Hz
90
1 kHz
0
5 kHz
0
10 kHz
0
Performing a Calibration Check
Measured
Deviation %
Value (A)
(AUX)
Measured
Deviation %
Value (V)
(NORMAL)
Maintenance
90-Day
Spec. (%)
0.161%
0.161%
0.161%
0.161%
0.161%
0.241%
0.541%
0.541%
90-Day Spec. (%)
0.101%
0.101%
0.038%
0.048%
0.048%
7
7-19