Setting A Dual Ac Voltage Output - Fluke 5500A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

5500A
Operator Manual

4-28. Setting a Dual AC Voltage Output

Tie the terminals NORMAL LO and AUX LO together at the UUT or at the
5500A, via the "LO"s softkey selection "tied."
The calibrator produces a dual ac voltage output by sourcing one ac voltage on the
NORMAL outputs and a second on the AUX terminals. See "Setting AC Voltage
Output" above for information on selecting an ac voltage output in dBm.
Complete the following procedure to set a dual ac voltage output. If you make an entry
error, press
Verify the applied voltage to the UUT does not exceed the
rating of the UUT insulation.
1. Press
2. Connect the UUT as described earlier in this chapter under "Connecting the
3. Set the UUT to measure dual ac voltage on the desired range.
4. Press the numeric keys and decimal point key to enter the desired voltage output at
5. Press a multiplier key, if necessary. For example, press
6. Press
7. The Control Display now shows the amplitude of your voltage entry. For example,
The AUX output is limited to 3.3 V rms for sinewaves, 6.6 V peak-to-peak
for squarewaves, 9.3 V peak-to-peak for triangle and truncated sinewaves.
8. Press the numeric keys and decimal point key to enter the desired voltage output at
9. Press a multiplier key, if necessary. For example, press
10. Press
11. The Control Display now shows the amplitude of your entries for the NORMAL
4-30
one or more times to clear the display, then reenter the value.
to clear any output from the 5500A.
Calibrator to a UUT."
the NORMAL terminals (maximum six numeric keys). For example, 123.456.
.
123.456 mV (below).
the AUX terminals (maximum six numeric keys). For example, 234.567.
.
terminals (upper reading) and AUX terminals (lower reading) (below is typical).
Note
Caution
Note
.
.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents