Ac Voltage Amplitude Accuracy, Squarewave (Aux) - Fluke 5500A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

7-26. AC Voltage Amplitude Accuracy, Squarewave (AUX)

This verification test is optional. It is not necessary to guarantee the full
calibration of the instrument. However, it may be useful when
troubleshooting an instrument when other functions and/or parameters are
incorrect.
The AC Voltage Amplitude Accuracy, Squarewave (AUX) tests the amplitude accuracy
at the AUX terminals. For this test, use the Fluke 5790A, and an rms-responding meter.
For squarewaves, the measured value (in rms) should be exactly half the nominal peak-
to-peak value.
Nominal
Nominal
Value
Value
(V p-p)
(V p-p)
NORMAL
(AUX)
3 V
300 mV
3 V
300 mV
3 V
300 mV
3 V
300 mV
3 V
3 V
3 V
3 V
3 V
3 V
3 V
3 V
Note
Measured
Frequency
Value
(Hz)
(AUX V rms)
10 Hz
1 kHz
5 kHz
10 kHz
10 Hz
1 kHz
5 kHz
10 kHz
Performing a Calibration Check
Deviation %
1-Year Spec. (%)
1.350%
0.800%
6.100%
6.100%
1.350%
0.800%
6.100%
6.100%