1-39. AC Current (Non-Sine Wave) Specifications (cont)
Squarewave
[1]
Ranges
2.9 to 65.999 mA
66 to 659.999 mA
0.66 to 4.39999 A
4.4 to 22 A
[1]
All waveforms are peak-to-peak output ranges.
[2]
Uncertainty is stated in peak-to-peak. Amplitude is verified using an rms-responding DMM.

1-40. AC Current, Square Wave Characteristics (typical)

Range
I<4.4 A @ 400 Hz

1-41. AC Current, Triangle Wave Characteristics (typical)

Linearity to 400 Hz
0.3 % of p-p value, from 10 % to 90 % point
1-Year Absolute Uncertainty,
Frequency
+ (% of output + % of range)
%Output
0.01 to 10 Hz
5.0
10 to 45 Hz
0.25
45 to 1 kHz
0.25
1 to 10 kHz
5.0
0.01 to 10 Hz
5.0
10 to 45 Hz
0.25
45 to 1 kHz
0.25
1 to 10 kHz
5.0
10 to 45 Hz
5.0
45 to 1 kHz
0.5
1 to 5 kHz
5.0
45 to 500 Hz
2.0
500 to 1 kHz
5.0
Rise Time
Settling Time
25 s
40 s to 1 % of final value
Introduction and Specifications
tcal + 5 C,
[2]
%Range
0.5
0.5
0.25
0.5
0.5
0.5
0.5
1.0
1.0
0.5
1.0
0.5
1.0
<10 % for loads <100
Aberrations
<1 % of p-p value, with amplitude >50 % of range
Additional Specifications
Maximum
Current
Resolution
Two digits, e.g., 50 mA
Six digits on each range
Two digits
Six digits on each range
Two digits
Six digits on each range
Two digits on each range
Six digits on each range
Overshoot
1