Ac Voltage, Dc Offset Specifications; Ac Voltage, Square Wave Characteristics; Ac Voltage, Triangle Wave Characteristics (Typical); Ac Current (Sine Wave) Extended Bandwidth Specifications - Fluke 5500A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

1-34. AC Voltage, DC Offset Specifications

[1]
Range
(Normal Channel)
3.3 to 32.999 mV
33 to 329.999 mV
0.33 to 3.29999 V
3.3 to 32.9999 V
9.3 to 92.999 mV
93 to 929.999 mV
0.93 to 9.29999 V
9.3 to 92.9999 V
6.6 to 65.999 mV
66 to 659.999 mV
0.66 to 6.59999 V
6.6 to 65.9999 V
[1]
Offsets are not allowed on ranges above the highest range shown above.
[2]
The maximum offset value is determined by the difference between the peak value of the selected voltage output and the
allowable maximum peak signal. For example, a 10 V p-p square wave output has a peak value of 5 V, allowing a maximum
offset up to ± 50 V to not exceed the 55 V maximum peak signal. The maximum offset values shown above are for the minimum
outputs in each range.
[3]
For frequencies 0.01 to 10 Hz, and 500 kHz to 2 MHz, the offset uncertainty is 5 % of output, ± 1 % of the offset range.

1-35. AC Voltage, Square Wave Characteristics

Rise Time
Settling Time
@ 1 kHz
@ 1 kHz
Typical
Typical
<10 s to 1 %
<1 s
of final value
[1]
For duty cycles of 10.00 % to 90.00 %.

1-36. AC Voltage, Triangle Wave Characteristics (typical)

Linearity to 1 kHz
0.3 % of p-p value, from 10 % to 90 % point

1-37. AC Current (Sine Wave) Extended Bandwidth Specifications

Range
All current ranges, <330 mA
Max Peak
[2]
Offset Range
Sine Waves (rms)
0 to 50 mV
0 to 500 mV
800 mV
0 to 5 V
0 to 50 V
Triangle Waves and Truncated Sine Waves (p-p)
0 to 50 mV
0 to 500 mV
800 mV
0 to 5 V
0 to 50 V
Square Waves (p-p)
0 to 50 mV
0 to 500 mV
800 mV
0 to 5 V
0 to 50 V
Overshoot
@ 1 kHz
Duty Cycle Range
Typical
1 % to 99 %, <3.3 V p-p,
<2 %
0.01 Hz to 100 kHz
1-Year Absolute Uncertainty,
Frequency
± (% of output + % of range)
% Output
0.01 to 10 Hz
10 Hz to 10 kHz
Introduction and Specifications
1-Year Absolute Offset
Uncertainty, tcal ± 5 C
Signal
± (% Output (dc) + V)
80 mV
8 V
0.1 + 3300
55 V
0.1 + 33 mV
80 mV
8 V
0.1 + 9300
55 V
0.1 + 93 mV
80 mV
8 V
0.1 + 6600
55 V
0.1 + 66 mV
Duty Cycle Uncertainty
± (0.8 % of period +140 ns) for
frequencies >10 kHz; + (0.8 % of
period + 2 s) for frequencies
Aberrations
<1 % of p-p value, with amplitude >50 % of range
tcal ± 5 C,
% Range
5.0
0.5
(See AC Current (Sine Wave) Specifications)
Additional Specifications
[3]
0.1 + 33
0.1 + 330
0.1 + 93
0.1 + 930
0.1 + 66
0.1 + 660
[1]
10 kHz.
Maximum
Current Resolution
2 digits each range
1
1-29

Hide quick links: