Required Test Equipment
The equipment listed in Table 2 is necessary to do the performance verification tests and
calibration adjustment.
Table 2. Required Equipment for Verification and Calibration
Equipment
DC Calibrator
Digital Multimeter
HART Temperature Transmitter
Flexible Test Leads
Double Banana Jack Cable
Performance Verification Tests
To prevent possible electrical shock, fire, or personal injury, do
not perform the performance test or verification test procedures
unless the Product is fully assembled.
The performance tests verify the complete operation of the Product and measure the
accuracy of each function against Product specifications. If the Product fails a part of the
test, calibration adjustment or repair is necessary. See "Calibration Adjustment
Procedure".
The performance verification tests check the accuracy of each Product function against its
specifications. If the Product fails any of these tests, calibration adjustment or repair is
necessary.
Recommended Model
Fluke 5522A Multi-Product
Calibrator
Fluke 8508A
Rosemount 644
Fluke TL24
-
Warning
mA Loop Calibrator
Required Test Equipment
Minimum Specification
DC Voltage: 0 V to 30 V
Accuracy: ±0.002 % +0.5 mV
DC Current 0 mA to 24 mA
Accuracy: ±0.002 % +0.5 μA
DC Current: 0 mA to 26 mA
Accuracy: ±0.002 % +0.5 μA
-
-
-
7