10.5

The factory check list and factory test report

The factory check list and factory test report is designed for the ReVo and ReVo Pro cabinets
specifically. They document all quality related activities performed during the production through
testing and cleaning when the cabinet is ready for packing and shipment.
A filled in and reviewed check list and test report is shipped with each cabinet.
10.6

Test equipment needed to perform installation tests and service tests

The test equipment needed to perform installation and service test are:
Test
Leak test of HEPA filters
Downflow air velocity and setting of alarms
Inflow air velocity and setting of alarms
Test of alarms
Light test
Airflow pattern test
All equipment used should be of a suitable quality and should be calibrated where deemed
necessary in compliance with EN 12469.
10.7

Test information

For more detailed information on the individual tests please refer to the factory test report delivered
with the cabinet.
During installation and service testing, the values measured during the factory test are the target
values to be reached as close as possible.
From the factory, all parameters and functions has been set to ensure that all requirements are meet
and that the cabinet will protect the samples handled, the operator doing the work and the
environment in the surrounding room.
All tests should be done by authorized test technician approved by Labconco.
11

Options for ReVo and ReVo Pro cabinets

A number of optional components can be fitted to the ReVo and ReVo Pro cabinets.
The cabinets can be ordered with these options or the options can be retro-fitted by trained service
technicians.
The most common options are listed below. Contact your distributor or Labconco for availability of
other options.
11.1

LED light

LED lighting is standard. LED lighting provided with the cabinet is dimmable.
The use of LED light saves on energy versus traditional fluorescent light tubes.
Smoke generator and aerosol photometer or
particle counter and diluter
Thermo-anemometer
Thermo-anemometer
No equipment needed
Light meter
Smoke stick or similar