Connections; Assemble The Stand For The Revo Cabinets - Labconco ReVo User Manual

Class ii, type a2 microbiological safety cabinets

Advertisement

5.2

Connections

The wall socket to which the cabinet is connected must be fused by a minimum 10 A fuse. The wall
socket should be placed so accidental switch off is unlikely. A wall socket without a switch is
recommended. Wires must be live, neutral and earth.
Required fusing: Circuit breaker 10 A (13 A) or fuse T 10 A (T 13 A).
In addition, the applicable safety requirements of the local power Supply Company shall apply.
Caution
If an automatic relay for disconnection of power in case of electric fault is needed, it
must be installed in the supply, as it is not built-in.
Caution
Before connecting the power supply, it must be checked that the mains specifications
correspond to those stated on the type plate.
Caution
The safety cabinet is provided with a flexible power cord. The connection may be
installed hard-wired or by means of a wall outlet with protective ground.
If a hard-wired connection is used, a main switch, which will cut off all poles to the unit, must be
used. This switch must be lockable both in the ON and OFF positions.
5.3

Assemble the stand for the ReVo cabinets

As default the ReVo is delivered without the stand mounted. ReVo Pro is always delivered with the
stand mounted.
All nuts and bolts for assembly of the stand and for mounting the stand to the cabinet are delivered
with the stands.
Start by assembling the stand and then place the cabinet on top and fasten with four bolts – one in
each corner.
Stand
Electrical stand
22