6.11

Interrupting an ongoing procedure

To interrupt a work process:
1.
Remove all samples from the working chamber and store them properly.
2.
Remove any accessories from the chamber, and clean and disinfect them.
3.
Clean and disinfect the working chamber surfaces, work tray, and the trough.
4.
Turn the cabinet either to standby mode or off mode.
6.12

Shut down of the ReVo and ReVo Pro cabinets

If the cabinet is not to be used or is to be stored for an extended period of time, it must be completely
decontaminated.
Warning
To shut the cabinet down, the sample chamber must be disinfected completely and
the plenum, including the filters, must be decontaminated using the appropriate and
authorized procedures.
1.
After the device has been decontaminated, close the window completely.
Disconnect the unit from the power source.
2.
6.13

Disposal of cabinets

All components except for the HEPA filters can be discarded after having been thoroughly cleaned
and decontaminated. The HEPA filters must be discarded in accordance with the applicable
international, national and local regulations for chemical or biological waste.
Warning
As this cabinet, can be used for processing and treating infectious substances, it must
be decontaminated prior to disposal, in accordance with acceptable standards and
procedures.
6.14

Trouble shooting

Item
Fault message
Ventilators does not
Cabinet
start
Light does not turn on
Light
with cabinet running
in normal mode
Front window
No movement
Power up error
System alarms
Low flow alarm
Flow 1 = downflow
Fault reaction
No reaction when trying to start
cabinet
Cabinet running without light in
the working chamber.
No movement
"Power up error" in display is
show to show that the cabinet
has been without power.
Visible "Low flow X" and audible
alarm on control panel
33
Action
Check:
Power supply to cabinet
Mains cable is in place
Call service technician
Check:
Change light tubes
Change ballast
Call service technician
Check:
Blockage of window
Call service technician
Press the ENTER button.
Check: