8.5

Replacement of laminator

The laminator is placed at the top of the working chamber just below the main downflow filter. To
remove the laminator, the two fixing screws, shown on the picture below, must be removed.
Pull on the two flaps located at the back edge of the laminator and pull downwards. Be careful not
to damage the main downflow filter located just above the laminator.
Install the replacement laminator and mount the fixing screws again.
8.6

Replacements and repairs

All replacement and repairs – other than the two described above – must be performed by a trained
service technician.
8.7

Disposal of cabinet

The entire cabinet except for the HEPA filters can be discarded after having been thoroughly cleaned
and decontaminated. The HEPA filters must be discarded in accordance with the applicable national
and local regulations for chemical and biological waste.
Warning
As this cabinet can be used for processing and treating infectious substances, it may
become contaminated. Prior to disposal, the entire cabinet with filters must be
decontaminated in accordance with acceptable standards and procedures.
Product name
Laminator ReVo and ReVo Pro 900
Laminator ReVo and ReVo Pro 1200
Laminator ReVo and ReVo Pro 1500
Laminator ReVo and ReVo Pro 1800
40
Catalogue No.
R3832500
R3832501
R3832502
R3832503