and 300 mm strait or elbowed duct diameters, either direct or by thimble, allowing simultaneous air
extract from the room.
Continuous operation of the safety cabinet helps to control dust levels and other airborne particulates
in the laboratory. If the safety cabinet is only operated when needed or put into reduced speed mode
(NightSmart™) to conserve energy, the balancing of laboratory room air must be taken into account.
If the room exhaust is balanced to include air exhausted through the ducted safety cabinet, this
cabinet must not be turned off.
The ReVo and ReVo Pro cabinets can exhaust in three common ways:
1.
Directly to the room in which they are installed.
2.
Directly to the room in which they are installed through double exhaust filters. Normally used
only on ReVo cabinets.
3.
Through thimble with anti-blowback to an external ventilation system.
4.5
Cabinet exhaust system accessories
The safety cabinet can be equipped with a range of exhaust system accessories as listed below:
4.5.1
Double exhaust HEPA filter (Not GS)
Normally factory fitted but can be retro fitted if required.
Required in UK for recirculating Class II cabinets when working
with Cat 3 biohazard material. Available for all cabinet models
with 2 sizes available. One for 900 and 1200 cabinets and one for
1500 and 1800 cabinets.
4.5.2
Double exhaust Carbon filter (Not GS)
Normally factory fitted but can be retro fitted if required.
Required if working with organic / inorganic solvents that need to
be removed from exhaust air. Available for all cabinet models with
2 sizes available. One for 900 and 1200 cabinets and one for
1500 and 1800 cabinets.
Product name
Double exhaust 900/1200
Double exhaust 1500/1800
Product name
Carbon filter 900/1200
Carbon filter 1500/1800
18
Catalogue No.
R2502500
Contact Labconco
Catalogue No.
Contact Labconco
Contact Labconco