• Precision calibrated syringe detection is designed for optimal fluid injection into an
empty fluid path.
• Start waste volumes are sufficient to wet and flush the channels prior to formula-
tion.
Follow the steps below to prime the system. If not using priming, proceed to
6.3.6 Load syringes and collection tubes, on page
Step
1
2
3
4
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA
Action
Insert the waste collection tube into the left clip of the sample arm.
Fill syringes with the priming solvents. At least 6 mm (¼ inch) of fluid must be
added to each syringe.
Open the lid of the instrument.
Rotate the rotating block upwards so that the underside of the cartridge is
visible.
Result:
6.3.5 Prepare the cartridge
94.
6 Operation
6.3 Run a procedure
Section
91