Step
10
Proceed to the next section to start the formulation. It is recommended to start the
formulation as soon as possible after loading the syringes.
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA
Action
Close the lid of the instrument.
CAUTION
Do not wait longer than 3 minutes between inserting the syringes
and starting the formulation. Leaving the heating block on with the
syringes inserted can cause ignitable vapor to form inside the
instrument.
6.3 Run a procedure
6.3.6 Load syringes and collection tubes
6 Operation
100