6.3.3

Optional: Preheat the system

Introduction
The Ignite and Ignite+ systems offer optional heating, for procedures that use pre-
heated liquids. The heating block preheats the syringes and holds the temperature of
the syringes in the C and R channels above ambient temperature for the duration of
the run.
The table below gives the limitations on temperature for each of the instruments.
Instrument
Ignite
Ignite+
Preheat the instrument
Before you begin, make sure that the heating insert or inserts are correctly sized for the
syringes used in the run. See
Note:
Step
1
2
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA
IMPORTANT
While C and R syringes are heated independently, they are only
able to maintain a maximum offset of 30°C. For example, if the R
syringe temperature is set to 75°C and the C syringe temperature
is set to 25°C, the actual C syringe temperature will reach 45°C.
Section 3.5.1 Heating insert, on page 38
Only C and R channels can be used with heating.
Action
Open the lid of the instrument.
Rotate the rotating block upwards so that the underside of the cartridge is
visible.

6.3.3 Optional: Preheat the system

Temperature
Up to 75°C
Up to 65°C
6 Operation
6.3 Run a procedure
for compatibility.
83