Specification-Type 576
Vertical Step
3%
4%
5%
4%
Generator
Horizontal Step
3%
4%
5%
4%
Generator
Deflection Factors
Vertical
Collector Current
1 !-LA/division to 2 A/division in
1-2-5 sequence.
Emitter Current
1 nA/division to 2 rnA/division in
1-2-5 sequence.
Step Generator
1 step/division.
Horizontal
Collector Volts
50 mV /division to 200 V /division
in 1 ·2-5 sequence
Base Volts
50 mV /division to 2 V/division in
1-2·5 sequence.
Input I mped-
At least 100 MQ with HORIZON-
a nee
TAL switch set to 50 mV, 100 mV
and 200 mV BASE; 1 MQ within
2% with switch set to .5 V, 1 V and
2V.
Step Generator
1 step/division
Maximum Displayed
Noise
1% or less, or the following depend-
ing on setting of MAX PEAK
VOLTS switch:
15
75
350
1500
Vertical
COLLECTOR
1!-LA
1!-LA
2 !-LA
5!-LA
EMITTER
1 nA
1 nA
2 nA
5nA
Horizontal
COLLECTOR
5mV
5mV
20 mV
200mV
BASE
5mV
5mV
5mV
5mV
Calibration Check
With DISPLAY OFFSET Selector
switch set to NORM (OFF), spot is
deflected 10 divisions both vert-
ically and horizontally within 1.5%
whenever the CAL button is pres-
sed.
With DISPLAY OFFSET Selector
switch set to X10 MAGNIFIER
(either axis) the calibration spot is
within 0.5% of zero spot (previous
ly set to CRT graticule center)
when CAL button is pressed.
1-4
Vertical and Horiz-
Coarse positioning in 5 division in-
ontal Position Controls crements within 0.1 division; con-
tinuous fine positioning over at
least 5 divisions for each coarse pos-
ition.
Display Offset
Vertical or Horizontal offset of dis-
play centerline value up to 10 divis-
ions in 21 half division steps.
Display Positioning
Spot positioning with change in
Accuracy Using
POLARITY switch setting (using
POLARITY Switch
AC position as reference). within
0.1 division of:
Vertically
Horizontally
AC
Centered
Centered
+(NPN)
-5 divisions
-5 divisions
-(PNP)
+5 divisions
+5 divisions
CRT and Readout
CRT
Type
Electrostatic deflection.
Screen Size
Calibrated area of 10 divisions by
1 0 divisions; 12 usable divisions
horizontally ( 1 division equals 1
em).
Typical Accel-
4000V
lerating Po ten-
tial
Readouts
Automatic digitally lighted display.
Readout is automatically blanked if
readings would be outside the avail-
able ranges or would give erroneous
display.
PER VERT DIV
1 nAto 20 A calculated from VER-
TICAL switch setting, DISPLAY
OFFSET Selector switch setting
and MODE switch setting (or X 10
Vertical Interface Input).
PER HORIZ DIV
5 mV to 200 V calculated from
HORIZONTAL switch setting and
DISPLAY OFFSET Selector switch
setting
PER STEPS
5 nA to 2A and 5 mV to 20 V cal-
culated from AMPLITUDE switch
setting and .1X STEP MULT but-
ton position (or X10 Step Interface
Input).
©