DIGICAT 500i/550i, Functional Checks
Appendix A
Functional Checks
A.1

Locator Functional Check

Checking the function
Before any tests can be carried out it is vital to check the status of the unit, its batteries
and basic functionality.
The following list is used to achieve this.
1. Inspection
Once the general condition of the Locator is established the Audio Visual test can be
performed.
2. Audio / Visual display test
Upon depressing the trigger the Locator should test the display and speakers by illuminating
each segment in the bar display, the mode and function indicators and depth display, the
battery indicator light will illuminate throughout the display test. All LCDs must be operative
and an audible output must be heard.
• Casing
The casing should be free of significant damage.
• Labels
Body labels must be legible and intact. Display label must be
free of damage and tears.
The hatch must lock into place.
• Battery hatch
• Battery holder
All the battery contacts and springs on the holder must be free
of corrosion and the holder in good condition.
• Battery contacts
The battery contacts must be free of corrosion.