Appliance Data Storage, Overwriting - Smeg Tethys D60 User Manual

Washer disinfector
Table of Contents

Advertisement

Set
PARAMETER
Test cycle
(Parameter Introduced from the Main Firmware version
1.11.0.33) The parameter can be used by authorized
technicians only; it allows the display and the start of a
Test program, on the display "Pr t". This test program
performs a short washing cycle and activates all the
loads of the device. The cycle is defined to be used for
-t
machine verification after components replacement.
- Parameter value "--", you do not have access to the
cycle "Pr t"
- Parameter Value "on", the technician gets access to the
Test Program "Pr t".
In order not to run the Test cycle "Pr t": switch Off the
device to reset its condition.
In the following pictures some examples of "Set" parameters are featured.
fig. 26 – "Set - t" value can be "--" or "on". In this latter case a test cycle will be activated.
6.1

APPLIANCE DATA STORAGE, OVERWRITING

The appliance stores the data relating to the cycles performed and all the alarms which have occurred in its own
internal memory, on a PCB.
The data in the appliance's memory can be read using the TRACELOG software.
Overwriting of the data in the appliance's memory can be enabled or disabled using a setup parameter, Set: -b – ; for
details, refer to the SETUP MODE - "FUNCTIONS" section.
Parameter Set -b has two possible values:
- "0" meaning "overwriting disabled"
- "1" meaning "overwriting enabled" – 1 is the default value for the parameter; overwriting is permitted and no alarm
is triggered when the appliance's memory is full.
WARNING
If overwriting is disabled, parameter "-b=0", alarm AF:91 is triggered once the memory is full.
In this case the appliance is cut out and will not perform any more cycles until the data from the
memory are downloaded to free spaces; this can only be done using the TRACELOG software.
User Manual
DESCRIPTION
Tethys D60 - Tethys T60 - Tethys T45
POSSIBLE SUPERUSER ACTION
With set parameter
With set parameter
"-d=0"
Display Only
Display Only
Page 39 - 77
"-d=0"

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tethys t60Tethys t45

Table of Contents