"SET" PARAMETERS

Set on the display.
To access this item press the Start/Stop button.
The Set item contains a number of setting parameters.
User level operators can only display the parameters and cannot modify any "Set" parameters.
Set
PARAMETER
Detergent pump P1 dispensing rate.
Setting in ml/litre, with maximum threshold of 20
ml/litre.
-1
The default value of this parameter is 4ml/litre.
Maximum recommended quantity with recommended
detergents: 10ml/litre.
Increase and decrease with the Selection buttons.
Detergent pump P2 dispensing rate.
Setting in ml/litre, with maximum threshold of 20
ml/litre.
-2
The default value of this parameter is 2ml/litre.
Maximum recommended quantity with recommended
detergents: 10ml/litre.
Increase and decrease with the Selection buttons.
Detergent pump P3 dispensing rate.
Setting in ml/litre, with maximum threshold of 20
ml/litre.
-3
The default value of this parameter is 10ml/litre.
Maximum recommended quantity with recommended
detergents: 15ml/litre.
Increase and decrease with the Selection buttons.
Water hardness setting
Set the hardness in °f, in steps of 5. Range: 5-60°f. The
-4
default setting is 40°f.
Warning: Make sure that the value is actually the same
as the hardness of the water used.
Printer language
English by default, the language can be chosen from the
various options available
-6
German
dE
Demineralised water
The use of demineralised water can be activated or
-7
deactivated depending on whether or not the relative
connection has been made (Activated: dn, Deactivated: -
-)
Filter hours (Tethys D60 only)
The parameter indicates, to the nearest ten, the number
of absolute filter hours left before the maintenance
-8
message appears. The initial default setting is 500. (e.g.
the display shows "-8:50"). The value can be reset by an
authorised engineer.
User Manual
DESCRIPTION
French
Spanish
Italian
Fr
ES
It
Tethys D60 - Tethys T60 - Tethys T45
POSSIBLE SUPERUSER ACTION
With set parameter
"-d=0"
Display and
Modification
Display and
Modification
Display and
Modification
Display and
Modification
Display and
Modification
English
En
Display and
Modification
Display Only
With set parameter
"-d=0"
Display
Display
Display
Display and
Modification
Display and
Modification
Display and
Modification
Display Only
Page 37 - 77